Product Details:
Payment & Shipping Terms:
|
Model: | ASli | Temperature: | -70°C ~+150°C(200 °C Optional) |
---|---|---|---|
Humidity: | 20%~98%R.H (5%~98%R.H Optional) | Temp. Analytic Accuracy: | 0.1°C |
Temp. Distribution Uniformity: | +/-2.0°C | Humi.analytic Accuracy: | 0.1% R.H |
Humi.distribution Uniformity: | +/- 3.0% R.H | Temp. Rising Rate: | 3.0°C/min (customized) |
Temp.Falling Rate: | 1.2°C/min(customized) | Material: | SUS 304# Stainless Steel |
Compressor: | French Tecumseh, Germany | Power: | 220V / 380V |
High Light: | humidity and temperature controlled chamber,temperature and humidity test chamber |
(Humidity-freeze Thermal Cycling Test) High-low Temperature Humidity Test Chamber
Brief Introduction for PV Modules test standards
The designed usage time for solar cell module is about 20~30years , and the reliability test is the simulation of the earth's surface photovoltaic modules (crystalline silicon and amorphous silicon thin film, concentrating) design verification, can let the module long-term operating more than 20 years in a general climate , specification requires that solar cells module do test below :
Thermal cycle test;
Humidity-freeze test;
Damp Heat;
To confirm the solar cell can withstand high temperature and high humidity level with the minus temperature influence, as well as for repeated change caused by the temperature and thermal fatigue failure, in addition to determine solar cells resistant to moisture long-term penetration ability of the test equipment design and ability to meet the relevant standards below:
(IEC61646 , IEC61416 ,IEC61215 , UL1703 , IEC62108 , IEEE1513 ,IEC61730 etc ) .
No. | Test Name | Test Purposes | Equipment Name | Equipment Model | Main Technical Parameters | Brand |
1 | Damp Heat Test (Double 85 Test) |
To determine the ability of the module to withstand the effects of long-term penetration of humidity. | (Double 85 Test) High Temperature &High Humidity Test Chamber |
Test Requirement: Method:Put the modules into the chamber according national standard and IEC61215UL1703,Test Temperature 85 ± 2 ℃,Test Humidity 85% ± 5%,Test Time 1000h or 3000H. |
ASLi | |
2 | Humidity-Freeze Test | The purpose of this test is to determine the ability of the module to withstand the effects of high temperature and humidity followed by sub-zero temperatures. This is not a thermal shock test |
(Humidity-freeze Thermal Cycling Test) High-low Temperature Humidity Test Chamber |
Test Requirement: Method:Put the modules into the chamber accoring to national standard andIEC61215UL1703,Test Temperature 85 ± 2 ℃,Test Humidity 85% ± 5%,Test time 20h ,then test under -40 ℃ for0.5h ,10times cycling ;which the temperature rising and falling speed requires:100℃or 200℃/H . |
ASLi | |
3 | Thermal Cycling Test | To determine the ability of the module to withstand thermal mismatch, fatigue and other stresses caused by repeated changes of temperature. |
High-Low Temperature Cycling Chamber | Test Requirement: Method:Put the specimen into the chamber according to national standard and IEC61215UL1703,Test Temperature-40 ℃,Test time 10M ,then under 85 ± 2 ℃for 10M ,continuously 200 times cycle,in which temperature rise/fall speed requires 100℃/H . |
ASLi | |
4 | UV Preconditioning Test | To precondition the module with ultra-violet (UV) radiation before the thermal cycle/ humidity freeze tests to identify those materials and adhesive bonds that are susceptible to UV degradation | UV Aging Test Chamber | Test Requirement: Method: Put the specimens into the UV aging chamber according national standard and IEC61215UL1703,using 340nm wavelength UV for component and raw materialradiation. |
ASLi | |
5 | Corrosion Test | To determine the module ability to withstand corrosion | Salt Spray Test Chamber | Test Requirement: Method: Put the specimens in the salt spray chamber according to national standard and IEC61215UL1703:Temeprature setting : saturated air barrel: 47℃(+/-)1℃ ; Testing room:35℃(+/-)1℃; Angle 35°. |
ASLi | |
6 | Hot Spot Endurance Test | To determine the ability of the module to withstand hot-spot heating effects, for example solder melting or deterioration of the encapsulation | Xenon Light Aging Test Chamber | Test Requirement: Method: Put the specimen into the xenon aging test chamber accoring to national standard and IEC61215UL1703:Temperature +25℃~+85℃;Wave Range:295-780nm;Humidity 45%~98% ; Rain Pressure 80~120KPA ; Testing Time 300H . |
ASLi |
Contact Person: Doris
Tel: 0086-13602389184
-70 to +150 Degree Climate Temperature Humidity Test Chamber Price
Walk-In Climate Test Chamber For Low High Temperature And Humidity Testing
HL-225-F High And Low Temperature Cycle Test Chamber Has Cyclical System
Walk-in Temperature Humidity Testing Room, Environmental Walk-in Chamber
Customized Walk In Climatic Test Chamber Environmental Test Equipment
20% - 98% Rh Walk In Climatic Stability Chamber Electronic For Auto Spare Parts
Sunshine Simulation Uv / Xenon Aging Room Walk In Stability Chamber For Color Fastness Test
Vertical Type Sine Shock Vibration Temperature Humidity Test Chamber
Electromagnetic Horizontal X / Y Axle vibration Test System 20 Kn Exiting Force
350000N Max Sine Force Vibration Measuring Instruments / Vibration Analysis Equipment
5 - 2000hz Vibration Test System 20kn Force With Vertical Horizontal Bench